ZL-1057 Dielectric Constant And Dielectric Loss Test System

Products

ZL-1057 Dielectric Constant And Dielectric Loss Test System

This high-frequency test system for insulating materials is composed of S916 test fixture, high-frequency Q meter, built-in data acquisition & automatic tanδ measurement software module, and HLK-1 inductors.

1.Introduction

This high-frequency test system for insulating materials is composed of S916 test fixture, high-frequency Q meter, built-in data acquisition & automatic tanδ measurement software module, and HLK-1 inductors.

It is designed and manufactured in compliance with GB/T 1409-2006, GB/T 1693-2007, ASTM D150 and IEC 60250 standards.

The system provides an optimal solution for automatic testing of high-frequency dielectric loss tangent (tanδ) and dielectric constant (ε) of insulating materials.

A parallel-plate capacitor fixture is used to clamp test samples, matched with Q meter as the reading unit. The dielectric loss tangent is calculated by formula based on Q value difference (sample loaded / unloaded) and sample thickness reading.

The digital Q meter can automatically compute and display dielectric constant (ε) and dielectric loss (tanδ).

 

2.System Composition

2.1 Main Unit: High Frequency Q Meter

Function Item Specifications
DDS Digital Synthesized Signal Source Frequency Range 10KHZ–70MHz
Frequency Coverage Ratio of Signal Source 7000:1
Signal Source Frequency Accuracy (6 significant digits) 3×10⁻⁵ ± 1 digit
Q Factor Measuring Range 1~1000, auto / manual range switching
Q Factor Resolution 4 significant digits, resolution: 0.1
Q Factor Measuring Error <5%
Inductance Measuring Range (4 significant digits, resolution: 0.1nH) 1nH ~ 8.4H, resolution: 0.1nH
Inductance Measuring Error <5%
Tunable Main Capacitance 30–550pF
Direct Capacitance Measuring Range 1pF ~ 2.5μF
Tunable Capacitance Error / Resolution ±1 pF or <1% / 0.1pF
Resonance Point Search Function Automatic Scanning
Preset Qualified Q Range 5–1000 with audible & visual alarm prompt
Q Range Switch Mode Automatic / Manual
Parameters Displayed on LCD Screen F, L, C, Q, Lt, Ct, tn, etc.
Auto Deduction for Residual Inductance of Unit & Test Leads (Exclusive Design) Available
Direct Reading for Large Capacitance (Exclusive Design) Max measurable value: 2.5μF
Measurement Precision of Dielectric Loss Factor 0.05%
Measurement Precision of Dielectric Constant 0.1%
Testable Sample Thickness 0.1mm – 10mm
USB Communication Interface Supported

 

2.2 S916 Digital Test Fixture for Dielectric Constant εr & Dissipation Factor tanδ

  •  Solid Samples: Standard test diameter Φ38mm; adjustable thickness, max. thickness <14mm
  •  Liquid Samples: Electrode diameter Φ38mm; optional liquid cup, inner diameter Φ48mm, depth 7mm(Optional)

 

2.3 HLK-1 Inductor Set

No Inductance Inductance% Min.Q Factor Distributed Capacitance Resonance Frequency Range (MHz) Applicable Test Frequency
1 0.1μH ±0.05μH 200 5pF 20~70 50MHz
2 0.5μH ±0.05μH 200 5pF 10~37 15MHz
3 2.5μH ±5% 200 5pF 4.6~17.4 10MHz
4 10μH ±5% 200 6pF 2.3~8.6 5MHz
5 50μH ±5% 200 6pF 1~3.75 1.5MHz
6 115μH ±5% 200 6pF 0.75~2.64 1MHz
7 1mH ±5% 150 8pF 0.23~0.84 0.5MHz
8 5mH ±5% 130 8pF 0.1~0.33 0.25MHz
9 14mH ±5% 90 8pF 0.072~0.26 0.1MHz