ZL-3320 sulfur nitrogen tester (one) is based on chemiluminescence principle and ultraviolet fluorescence fluorescence principle and computer technology combined research and development of a new generation of precision analysis instrument. Suitable for determination of naphtha, distillates, engine fuels and other petroleum products. Used for the determination of naphtha, distillates, engine fuels and other petroleum products.

ZL-3320 sulfur nitrogen tester (one) is based on chemiluminescence principle and ultraviolet fluorescence fluorescence principle and computer technology combined research and development of a new generation of precision analysis instrument. Suitable for determination of naphtha, distillates, engine fuels and other petroleum products. Used for the determination of naphtha, distillates, engine fuels and other petroleum products.

Execution standard

Applicable standards: SH/T 0657, ASTM D4629, SH/T 0689, ASTM D5453, GB/T11060.8

 

The instrument features

1.Total nitrogen content was determined by chemiluminescence method and total sulfur content was determined by ULTRAVIOLET fluorescence method.

2, improve the ability to resist impurities interference, avoid the electric quantity method of the titration tank tedious operation and therefore bring unstable factors, so that the sensitivity of the instrument is greatly improved.

3, the key parts of the system adopt imported devices, so that the performance of the whole machine has a reliable guarantee.

4, software intuitive easy to learn, standard curve and results automatically save, never lost data

 

Technical parameters

Sample types

Liquid, solid and gas

Determination method

Chemiluminescence (N), ULTRAVIOLET fluorescence (S)

Sample quantity

Solid sample: 1-20mg liquid sample: 5-20μL Gas sample: 1-5mL

Measuring range

Liquid

0.2mg/L-10000mg/L

 

 

Solid

1.0ppm – 10000ppm

Gas

1.0mg/m³-1.0000mg/m³

 

 

Accuracy of measurement

Fluorescent sulfur meter

Chemiluminescence nitrogen analyzer

Sulfur nitrogen one

Injection volume (μL)

RSD(%)

0.2

0.1

0.2/0.1

20

25

5

5

5

10

10

50

50

50

10

5

100

100

100

10

3

5000

5000

5000

10

3

Temperature control range

800 ℃ to 1000 ℃

The temperature control accuracytemperature control accuracy

±1℃

Air supply requirements

High purity argon: purity above 99.995% high purity oxygen: purity above 99.99%

Working power supply

AC220V±10%  50Hz(Exit voltage is optional)

Work rate

3000 W

Overall dimensions

Main machine: 510(W)×460(D)×410(H)mm Temperature control: 550(W)×460(D)×440(H)mm

The weight of the

Host: 20kg Temperature control: 40kg

 