ZL-1030 Voltage Breakdown Tester

Products

ZL-1030 Voltage Breakdown Tester

  Mainly applicable to solid insulating materials such as: plastics, films, resins, mica, ceramics, silica gel, glass, insulating varnish, etc. In air or liquid medium, measuring power frequency (48~62Hz) or corresponding DC voltage breakdown strength and Withstand voltage time. 

Applicable to continuous uniform boost or step-up boost, apply AC / DC voltage to the sample until breakdown, measure the breakdown voltage value, calculate the breakdown strength of the sample; use the rapid boost method to put the voltage Raise to the specified value, and the sample will not break down for a certain period of time. The specified value is the value of the withstand voltage of the sample.

Application:

  Mainly applicable to solid insulating materials such as: plastics, films, resins, mica, ceramics, silica gel, glass, insulating varnish, etc. In air or liquid medium, measuring power frequency (48~62Hz) or corresponding DC voltage breakdown strength and Withstand voltage time. 

Applicable to continuous uniform boost or step-up boost, apply AC / DC voltage to the sample until breakdown, measure the breakdown voltage value, calculate the breakdown strength of the sample; use the rapid boost method to put the voltage Raise to the specified value, and the sample will not break down for a certain period of time. The specified value is the value of the withstand voltage of the sample.

 

Standards compliant:

It complies with IEC60243-1 “Test method for electrical strength of solid insulating materials – Test under power frequency”, IEC 60243-2 “Test method for electrical strength of solid insulating materials – Corresponding to DC voltage test” and ASTM D149 test standard requirements.

 

Technical Parameters:

  1. Test voltage: AC 0~50 kV continuously adjustable;DC 0~50 kV continuously adjustable
  2. Transformer capacity: 10 KVA
  3. Breakdown voltage: 0~50kV
  4. Boost speed: 6 positions in 50, 100, 200, 500, 1000, 2000 (KV/S)
  5. Boost speed regulation: for automatic electric speed regulation
  6. Voltage measurement accuracy: ≤ 2%
  7. Leakage current selection: 1~20 mA
  8. Test method: DC test, AC test
  9. Boost mode: automatic program-controlled boost
  10. Power frequency power frequency: 50Hz sine wave
  11. Power supply: AC220V 50/60Hz
  12. Dimensions: L 1400* W 1040* H 1500 mm
  13. Weight: 200KG