• Dongguan instrument welcome you, we will be good quality, bold innovation, welcome new and old customers to visit and guidance, we will sincerely service for you!
86-135 3248 7540
DongGuan ZhongLi Instrument Technology Co., Ltd.


》》Home > Products > Environmental testing machine > Environmental testing machine

ZL-6037 PCT High Pressure Accelerated Aging Test Chamber
  • 【Product name】:ZL-6037 PCT High Pressure Accelerated Aging Test Chamber
  • 【Specifications】:ZL-6037
  • 【Range of use】:The high-pressure accelerated life test device is also called the high-pressure accelerated aging tester and the saturated high-pressure accelerated aging tester. It is widely used in the testing of the sealing performance of defense, aerospace, automotive parts, electronic parts, plastics, resins, magnet industries, pharmaceutical circuit boards, multilayer circuit boards, IC, LCD, magnets, lighting, lighting products and other products, related to The accelerated life test of the product is used in the design section of the product to quickly expose the defects and weaknesses of the product. Test the wear resistance and air tightness of its products.
  • 【Product Brief】:PCT High Pressure Accelerated Aging Test Chamber The high-pressure accelerated life test device is also called the high-pressure accelerated aging tester and the saturated high-pressure accelerated aging tester. It is widely used in the tes
Introduction

PCT High Pressure Accelerated Aging Test Chamber

The high-pressure accelerated life test device is also called the high-pressure accelerated aging tester and the

saturated high-pressure accelerated aging tester. It is widely used in the testing of the sealing performance of

defense, aerospace, automotive parts, electronic parts, plastics, resins, magnet industries, pharmaceutical circuit

boards, multilayer circuit boards, IC, LCD, magnets, lighting, lighting products and other products, related to The accelerated life test of the product is used in the design section of the product to quickly expose the defects and

weaknesses of the product. Test the wear resistance and air tightness of its products.

 

Standard: GB/T 29309-2012, IEC62108

 

Test chamber material:

Temperature range

RT-132℃

Test box size

350 mm x L500 mm), round test box

Overall dimensions

1150x 960 x 1700 mm (W * D * H) vertical

Inner barrel material

stainless steel plate material (SUS# 304 5 mm)

Outer barrel material

cold plate paint

Insulation material

rock wool and rigid polyurethane foam insulation

Steam generator heating tube

finned heat pipe-shaped seamless steel tube electric heater (platinum plating on the surface, anti-corrosion)

 

1.Control system:

a. Use Japanese-made RKC microcomputer to control the saturated steam temperature (using PT-100 platinum temperature sensor).

b. The time controller adopts LED display.

c. Use pointer to display pressure gauge.

2. Mechanical structure:

a. Round inner box, stainless steel round test inner box structure, in line with industrial safety container standards.

b. The patented packing design makes the door and the box more closely integrated, which is completely different from the traditional extrusion type, which can extend the packing life.

c. Critical point LIMIT mode automatic safety protection, abnormal reason and fault indicator display.

3. Security protection:

a. Imported high temperature resistant sealed solenoid valve adopts double loop structure to ensure no pressure leakage.

b. The whole machine is equipped with over-pressure protection, over-temperature protection, one-key pressure relief, manual pressure relief multiple safety protection devices, to ensure the user's use and safety to the greatest extent.

c. Back pressure door lock device, the door of the test chamber cannot be opened when there is pressure inside the test chamber.

4. Other accessory accessories

4.1 1 set of test racks

4.2 Sample tray

5. Power supply system:

5.1 The fluctuation of the system power supply shall not exceed ±10

5.2 Power supply: single-phase 220V 20A 50/60Hz

6. Environment & Facilities:

6.1 Allowable working environment temperature 5℃~30℃

6.2 Experimental water: pure water or distilled water